A Hardware-Software System for Accurate Segmentation of Phonocardiogram Signal

نویسندگان

1 Department of Medical Physics and Biomedical Engineering, School of Medicine, Shiraz University of Medical Sciences, Shiraz, Iran

2 Ionizing and Non-ionizing Radiation Protection Research Center (INIRPRC), Shiraz University of Medical Sciences, Shiraz, Iran

3 Novin Iran Specialized Clinic, Shiraz, Iran

4 School of Medicine, Shiraz University of Medical Sciences, Shiraz, Iran

5 Student Research Committee, Shiraz University of Medical Sciences, Shiraz, Iran

6 Faculty of Electrical Engineering, Shiraz University of Technology, Shiraz, Iran

doi
10.31661/jbpe.v0i0.2104-1301
چکیده

Background: Phonocardiogram (PCG) signal provides valuable information for diagnosing heart diseases. However, its applications in quantitative analyses of heart function are limited because the interpretation of this signal is difficult. A key step in quantitative PCG is the identification of the first and second sounds (S1 and S2) in this signal.Objective: This study aims to develop a hardware-software system for synchronized acquisition of two signals electrocardiogram (ECG) and PCG and to segment the recorded PCG signal via the information provided in the acquired ECG signal.Material and Methods: In this analytical study, we developed a hardware-software system for real-time identification of the first and second heart sounds in the PCG signal. A portable device to capture synchronized ECG and PCG signals was developed. Wavelet de-noising technique was used to remove noise from the signal. Finally, by fusing the information provided by the ECG signal (R-peaks and T-end) into a hidden Markov model (HMM), the first and second heart sounds were identified in the PCG signal.Results: ECG and PCG signals from 15 healthy adults were acquired and analyzed using the developed system. The average accuracy of the system in correctly detecting the heart sounds was 95.6% for S1 and 93.4% for S2.  Conclusion: The presented system is cost-effective, user-friendly, and accurate in identifying S1 and S2 in PCG signals. Therefore, it might be effective in quantitative PCG and diagnosing heart diseases.