Capillarys 3 Octa®: Analytical Performance Assessment for HbA1c Quantification

نویسندگان

1 Laboratory of Biochemistry and Toxicology University Hospital of Monastir, Tunisia. Faculty of Pharmacy, University of Monastir, Monastir, Tunisia.

2 Laboratory of Biochemistry and Toxicology University Hospital of Monastir, Tunisia. Faculty of Pharmacy, University of Monastir, Monastir, Tunisia.

3 Laboratory of Biochemistry and Toxicology University Hospital of Monastir, Tunisia. Faculty of Pharmacy, University of Monastir, Monastir, Tunisia.

4 Laboratory of Biochemistry and Toxicology University Hospital of Monastir, Tunisia. Faculty of Pharmacy, University of Monastir, Monastir, Tunisia.

5 Laboratory of Biochemistry and Toxicology University Hospital of Monastir, Tunisia. Faculty of Pharmacy, University of Monastir, Monastir, Tunisia.

doi
10.18502/ijdo.v14i2.9453
چکیده

Objective: Determination of HbA1c level is a precious indicator for therapeutic follow-up of patients with type 1 or type 2 diabetes. Our study aimed to evaluate the analytical characteristics of Capillarys 3 Octa® HbA1c measurement by capillary electrophoresis. Materials and Methods: Our study involved 265 venous whole blood specimens repeatability, intermediate fidelity, accuracy, linearity, ...