Agreement between whole-blood CRP measurement using Humacount 5DCRP and serum CRP measurement by immunoturbidimetry (Abbott Alinity CI Series): A comparative study
نویسندگان
doi
10.48309/jmpcr.2025.506212.1625چکیده
C-reactive protein (CRP), an acute-phase protein in response to infection, inflammation, and tissue damage, is useful for both diagnosis and monitoring. Nowadays, serum CRP with immunoturbidimetry is a popular method for measuring CRP. This method requires more time for preparation and larger sample volumes. Humacount 5D CRP , a hematology analyzer, offers faster, single-run Complete Blood Count and CRP tests, reducing turn-around-time (TAT). However, this new method should be validated and compared with the established method. The objective of this study was to evaluate the agreement between whole-blood Humacount 5D CRP and serum CRP level and immunoturbidimetry method (Abbott Alinity). The CRP levels of 65 patients were each examined using whole-blood samples in Humacount 5D CRP and serum samples in Abbott Alinity. The correlation of CRP values was evaluated using Spearman test. The reliability test used Cohen Kappa agreement coefficient. Bland-Altman curve test analysis was used to examine the difference in limit of agreement of CRP levels between those analyzers. Spearman test showed a very strong positive correlation (R=0.956, p<0.001). Kappa agreement test analysis revealed almost perfect agreement (R = 0.938, p < 0.001). Bland-Altman curve analysis showed a mean difference of -2.59 (95% CI -12.34 - 7.16), where of the 65 samples tested, only four samples had a difference in CRP levels outside the limit of agreement value. The whole blood CRP with Humacount 5D CRP is strongly agreed with CRP with Abbott Alinity. It can serve as a reliable CRP testing, provide rapid results to support faster clinical decision-making, and improve patient management.