Dosimetry of Three Cone Beam Computerized Tomography Scanners at Different Fields of View in Terms of Various Head and Neck Organs

نویسندگان
doi
10.5812/iranjradiol.34220
چکیده

Background: Marketing new radiography devices necessitates documenting their absorbed X-ray doses. Since the current literature lacks studies on new devices, we assessed the doses of two new devices that had not previously been assessed. Objectives: The new devices were compared to the Promax three dimensional (3D) scanner at two fields of view (FOV) in nine critical head and neck tissues and organs. Results: The average absorbed doses, respectively, for the large and small FOVs were 17.19 and 28.89 mGy in the Promax 3D, 19.25 and 35.46 mGy in the NewTom VGi, and 18.85 and 30.63 mGy in the NewTom 5G. The absorbed doses related to the FOVs were not significantly different (P value = 0.1930). However, the effective doses were significantly greater at the smaller FOVs / higher resolutions (P = 0.0039). The doses of the three devices were not significantly different (P = 0.8944). The difference among the nine organs/tissues was significant (Kruskal-Wallis P=0.0000). Conclusion: The absorbed doses pertaining to the devices and the FOVs were not significantly different, although the organs/tissues absorbed considerably different doses.